The software programs we use are listed below.
1. Use of NIST MS Interpreter for EI and MS/MS Spectra
We routinely use NIST MS Interpreter to understand EI and MS/MS (APCI, electrospray) spectra. It is critical part of entering compounds into both our EI and MS/MS libraries. It quickly allows us to determine if all the major fragments in our spectrum are consistent with the proposed structure. In addition, it can also be used to correlate fragment ions in a model compound's spectrum to its substructure in the NIST and Wiley EI and MS/MS libraries and our in-house libraries.
It is supplied free of charge with the NIST Version 2.0 Software. I have included screenshots of the MS Interpreter and a poster presented by NIST:
ms_interpreter_screenshots (2 pages)
2. Identifying “Known Unknowns” Using ChemSpider and Automated MS/MS Structure Correlation
A poster session was given at the 2012 ASMS conference in Vancouver. The work was done as a collaborative effort between Agilent, ChemSpider (RSC), and Eastman Chemical Company. The approach is described below with a link to the poster session in PDF format:
- Accurate mass data acquired by LC/MS in data-dependent MS/MS mode
- Agilent MassHunter Qualitative Analysis software used to find compounds and generate molecular formulas using monoisotopic mass, isotope abundance, and isotope spacing, as well as look for matching formulas for each fragment ion and its neutral loss from the precursor
- Prototype Agilent MassHunter Molecular Structure Correlator software uses ChemSpider interface to obtain candidate structures for a target molecular formula and associated number of references for candidate structures
- Calculation of correlation scores to explain MS/MS fragmentation pattern for each candidate structure using a “systematic bond-breaking” approach
- Results sorted by number of references and/or correlation score to evaluate structure candidates for identification of components
asms_2012_poster_agilent_chemspider
3. Identifying “Known Unknowns” Using SciFinder
Similar to ChemSpider approach, but does not let one search by accurate mass m/z values with a window. Only molecular formulae are employed. However, many other capabilities such as keyword and partial structure (substructure) can be utilized. Also, many more structures are found in the CAS Registry versus ChemSpider.