The Thermo GC Orbitrap is a very powerful tool for the identification of unknowns easily switching from EI to CI modes. In most cases, useful results will be obtained library searching Orbitrap EI spectra against commercial EI reference libraries. However, users need to be aware of the differences in their EI orbitrap spectra that can be quite confusing. The spectra in commercial libraries are obtained on quadrupole, magnetic, and possibly TOF analyzers which are much different than Orbitraps or Iontraps.
From my experiences, the library search match factor will be ~80%. Thus, the user will normally obtain useful search results. However, smaller MW compounds, TMS derivatives, and other classes of compounds can be yield much lower match factors and somewhat confusing spectra. This is due to many different reasons found in my work which I have detailed in the following three links:
Figure Summarizing Studies 1-3
Study 1: NIST Manual Search Comparing 480 EI Spectra with Example Spectra
Study 2: Artifact Ions Noted in EI Spectra of TMS (Trimethylsilyl) Derivatives
Study 3: NIST Automatic Search Comparing 3,147 EI Spectra
Similar differences observed at Eastman have been noted by others in the literature for EI analyses and References 1-3 are noted at the bottom of this post. My studies are all in EI mode. However, artifact ions in tandem (MSMS) product ions are also noted from the reaction with water in the orbitrap analyzer (Reference 4).


Thermo and Personal User's Libraries:
Thermo has an on-going effort to create a database of EI GC-Orbitrap spectra which they supply to users. A user should also consider adding spectra to their personal NIST library noting the origin of the spectrum.
Literature References of Differences
- Gas-Phase Chemistry in the GC Orbitrap Mass Spectrometer
- GC-Ortitrap Combination Can Give Specral Alteration
- Evaluation of Performance and Validity Limits of Gas Chromatography Electron Ionization-Orbitrap Detector for Fatty Acid Methyl Esters Analyses
- Unexpected Peaks in Tandem Mass Spectra Due to Reaction of Product Ions with Residual Water in Mass Spectrometer Collision Cells